Pressure Measurement
TTC PRESSURE
MEASUREMENT LAB #X
OBJECTIVES
1. Examine a strain gauge pressure
transducer.
2. Calibrate a pressure switch and
determine its dead band.
3. Calibrate a pressure transmitter and
test its operation.
THEORY
Pressure is a physical property which
can be represented in units of force per unit area (kPag, psig, etc.), or as
head of a column of known liquid (mm of mercury, inches of water, etc.). In order to properly calibrate pressure
transducers and pressure transmitters, pressure sources or pressure reference
instruments must be at least 5 times more accurate than the instruments being
calibrated.
The most fundamental pressure
reference source is obtained from dead weight testers and manometers. Accurate mechanical master gauges and
electronic pressure calibrators can also be used as secondary pressure
standards for calibration purposes.
EQUIPMENT
- Strain Gauge Pressure Transducer
(Viatran Model 108 )
- Pressure Switch (UE 555)
- Differential Pressure Switch (UE
455)
- Pressure Transmitter (Foxboro 841GM)
- Decade Box
- Fluke 8060A DMM
PRE-LAB
Strain
Gauge Pressure Transducer (GP-50)
A strain gauge undergoes a change in
resistance when it is strained (stretched or compressed). This small change in
resistance can be detected and changed into a proportional mV signal by use of
a bridge circuit. The Viatran Model 108 Pressure Transducer uses four active
strain gauges mounted on a diaphragm which is deflected when exposed to a
pressure.
1. Sketch the hook-up for testing the
Viatran Model 108 given the following equipment:
·
5 VDC Power Supply
·
DMM
·
Pressure gauge and regulator
Pressure Switch Calibration (UE 555 and UE 455)
Pressure switches open and close a set
of contacts which may trigger alarms or operate equipment to control the
pressure. These contacts allow
connection in one of two ways: “Common” connected to “Normally Open”, or “Common” connected to “Normally Closed”.
The method of connection depends on the required use.
Fig A shows the pressure switch when
it is in it’s
“off
the shelf”
state.
Fig B shows the pressure switch when
it is in it’s
operating state (i.e. pressure alarm state).
2.
Sketch the hook-up necessary for
testing the UE 555 pressure switch given the following equipment:
·
Pressure gauge and regulator
·
DMM to test switch state (measuring
resistance: low resistance shows that
switch is closed, high resistance shows that switch is open)
3. Explain
how to adjust the trip point of the UE J402k pressure switch.
Pressure
Transmitter (Foxboro 841GM)
This transmitter is designed to
provide a 4 to 20 mA output signal for a 0 to 25 psig input pressure signal.
4. Sketch the hook-up necessary to
calibrate the Foxboro 841GM pressure transmitter using the following equipment:
·
The bench regulator and test gauge to
produce the input pressure
·
The bench 24 Volt power supply to
power the PT
·
A DMM to measure the current output
5. Explain how to calibrate this
transmitter for a range of 0 to 160 kPag.
6. What is the maximum loop resistance for
the pressure transmitter if powered by a 24V power supply?
7. How
can we use tp1 & tp2 to calibrate the pressure transmitter?
PROCEDURE
Strain
Gauge Pressure Transducer (Viatran Model
108)
1. Connect up the Viatran Model 108
Pressure Transducer pressure transducer to the bench regulated pressure and
apply the pressures required to fill in table 1. Note, use the bench 5 V power supply to power
the transducer and the Fluke 8060A to measure the mV output. Note, the pressure gauge is not connected to
the regulated pressure, therefore connect one of the regulated outlets to this
gauge.
2. From the mV measured at pressure inputs
of 0 kPag and 120 kPag, use the input/output formula to calculate the mV that
you would expect to see at the other pressures in the table. Show your calculation for an input of 60
kPag.
3.
The calculated output values from
step 2 assume that this strain gauge is a perfectly linear device. Calculate the error between the measured and
calculated output. Record your results in table 1.
4. From your results, produce an error
graph. Comment on the linearity of this
pressure transducer. List possible
sources of error in your results.
Pressure Switch Calibration (UE 555 and UE 455)
5. Calibrate the UE 555 dual pressure
switch. Have one switch trip at 10 psig
rising, and the other switch trip at 20 psig rising. Use the bench regulator and Druck 145 DPI to
apply/indicate the pressure and the DMM to determine when the contacts
open/close. Use the Common and Normally
Open contacts. Note the state of the
contacts before and after the pressure trip.
Record the actual trip pressure in table 2.
6. Slowly decrease the pressure until the
contacts change back to their original state. Record this pressure and
calculate the dead band in psig. Record
your results in table 2.
7. Calibrate the UE 455 differential
pressure switch. Have one switch trip at
30 “W.C.
differential falling, and the other switch trip at 60 “W.C. differential falling.. Use the bench regulator and Druck 145 DPI to
apply/indicate the pressure and the DMM to determine when the contacts
open/close. Use the Common and Normally
Open contacts. Note the state of the
contacts before and after the pressure trip.
Record the actual trip pressure in table 3.
8. Slowly decrease the pressure until the
contacts change back to their original state. Record this pressure and
calculate the dead band in psig. Record
your results in table 3.
9. Comment on your observations and
explain the need for some dead band with pressure switches.
Pressure
Transmitter (Foxboro 841GM)
10. Calibrate this pressure transmitter for a
0 to 160 kPag input as per pre-lab question #4.
11 Check the calibration of this
transmitter, and calculate error values.
Record your results in table 4.
12. Calculate the calibrated accuracy of this
transmitter in % of span. Show your
accuracy calculation.
13. Compare the accuracy of this transmitter
to the manufacturers stated accuracy.
Explain possible sources of error in the calibration procedure.
14. Connect
a decade box in series with the output current loop. Initially, set the decade box resistance to
zero. Adjust the pressure input until
the loop current is exactly 20 mA. Next,
slowly increase the decade box resistance until the transmitter can no longer
maintain it’s
20 mA output (i.e. increase resistance until the current just starts to drop
below 20 mA.). Record the decade box
setting. Compare this resistance to the
expected maximum loop resistance and explain any discrepancies.
TTC RESULTS LAB #X
NAME: ____________________________ DATE: _______________
LAB
PARTNER:
_____________________________ CLASS:
_________________
Pre-Lab
Signature: ________________________
Strain
Gauge Pressure Transducer (GP-50)
Test
set-up (pre-lab Q#1):
Results:
Input
Pressure (kPag)
|
0
|
20
|
40
|
60
|
80
|
100
|
120
|
GP-50
Output
(mV)
|
|||||||
Calculated
Output
(mV)
|
|||||||
Linearity
Error
(mV)
|
Table 1
Sample
calculation of output at 60 kPag (step 2):
Error
Graph (step 4):
Discussion
and Observations (step 4):
Pressure
Switch Calibration (UE J402k)
Test
set-up (pre-lab Q#2):
Trip
adjustment (pre-lab Q#3):
Results:
Switch
1
|
Switch
2
|
|
Trip
Pressure (psig)
|
||
Reset
Pressure (psig)
|
||
Dead
Band (psig)
|
Table
2
Results:
Switch
1
|
Switch
2
|
|
Trip
Pressure (“W.C.)
|
||
Reset
Pressure (“W.C.)
|
||
Dead
Band (“W.C.)
|
Table
3
Discussion
and observations and need for dead band (step 9):
Pressure
Transmitter (Foxboro 841GM)
Test
set-up (pre-lab Q#4):
Calibration
procedure (pre-lab Q#5):
Maximum
loop resistance (pre-lab Q#6):
Use
of tp1 and tp2 (pre-lab Q#7):
Results:
Input
Pressure
(kPag)
|
Expected
Output
(mA)
|
Measured
Output
(mA)
|
Error
(mA)
|
0
|
|||
20
|
|||
40
|
|||
60
|
|||
80
|
|||
100
|
|||
120
|
|||
140
|
|||
160
|
Table
3
Accuracy
calculation (% of span)(step 12):
Discussion
and observations (step 11):
Maximum
loop resistance calculation and discussion (step 13):
Reference : NAIT Edmonton CA
0 ความคิดเห็น:
แสดงความคิดเห็น
สมัครสมาชิก ส่งความคิดเห็น [Atom]
<< หน้าแรก